IKV-1-1-2 type A vibration measuring channel

IKV-1-1-2 type A vibration measuring channel

Designed for measurement the amplitude of vibration acceleration in diagnostics systems;

Consists of DV-1 (type 00, 02) vibration transducer, AV 112 charge amplifier and TIK-BIS.3X3.0X0X safety barrier installed in the connecting box (the box and the barrier are delivered at the request of the Customer);

Supplied complete with anti-vibration cable;

Output signal type – voltage two-wire (IEPE)




Additional photos

Reviews

Related equipment

TIK-PLC
controllers
TIK-BIS
safety barriers

specifications

Metrological parameters

  • Conversion coefficient, mV*s2/m / Range of indications of amplitude of vibration acceleration, m/s2
    1000 / 0-2
    200 / 0-10
    66.66667 / 0-30
    20 / 0-100
    10 / 0-200
    6.666667 / 0-300
    4 / 0-500
    2 / 0-1000
  • Operating frequency range, Hz
    3-10 000

Interface

  • Output signal type
    voltage two-wire (IEPE)
  • Output DC voltage, V
    12±0.2
  • Maximum amplitude value of alternating voltage, measured by channel, V
    ~ 2
  • Power supply from explosion-proof current source, mA
    10±1

Explosion protection

  • Kind
    intrinsically safe circuit
  • Marking
    0Ex ia IIC T6...T1 Ga X
  • Connection
    through controller
    through safety barrier

Design parameters

  • Version
    IKV-1-1-2 type A (DV-1 type 00 vibration transducer or
    DV-1 type 02 vibration transducer, AV 112 charge amplifier,
    optional - TIK-BIS.3X3.0X0X safety barrier, connecting box)
  • Overall dimensions, mm
    Ø39.25x39.5 (DV-1 vibration transducer)
    98.5x27.5x59 (AV 112 charge amplifier)
  • Weight, kg, not more than
    0.2 (DV-1 vibration transducer)
    0.1 (AV 112 charge amplifier)
  • Mounting type
    screw M4-6g 3 PCs.
  • Protection class
    IP65/IP68 (DV-1 type 00 vibration transducer)
    IP65 (DV-1 type 02 vibration transducer)
    IP30 (AV 112 charge amplifier)

Operating parameters

  • Operating temperature range, ºС
    -60...+150 (DV-1 type 00/02 vibration transducer)
    -60...+70 (AV 112 charge amplifier)

Reliability parameters and manufacturer's guarantees

  • MTBF, hours, not less than
    40 000
  • Warranty service life, months
    24
  • Service life, years, not less than
    10
  • Installed device life, hours, not less than
    80 000
  • Time between verifications
    1 year
Service center