Sensors DVA141.121 and DVA141.221 is designed to measure the root mean square (RMS) value vibration velocity in the laboratory;

The stainless steel housings are integrated microelectromechanical acceleration sensor and conversion board;

Available sensors - ideal for training, experiments and simple tasks;

Type of output signal - 4...20 mA




Other photos

Comments

technical characteristics

Metrological parameters

  • Dynamic range of acceleration, m/s2
    not more than 160
  • Measuring range of RMS vibration velocity, mm/s
    0...10
    0...12,7
    0...20
    0...25
    0...25,4
    0...30
    0...40
    0...50
    0...50,8
    0...60
    0...80
    0…100
  • Operating frequency range, Hz
    2-1000
    5-1000
    10-1000
  • Electrical insulation resistance between the vibration transducer circuit and the shell, MOm, not less than
    40

Interface

  • Output signal type
    "current loop" 4...20 mA
  • Supply voltage range, V
    10...24
  • The polarity of the connection of the vibration transducer
    arbitrary

Explosion protection

  • Kind of intrinsically
    safe electrical circuit
  • Marking
    0Ex ia llC T5...T6 Ga X
  • Connection
    via TIK-PLC 112 controller
    via safety barrier

Construction parameters

  • Execution
    with РС-4 connector on the body
  • Overall dimensions, mm
    Ø39,25х60 (DVA141.121)
    Ø28х60 (DVA141.221)
  • Weight, kg, not more than
    0,17
  • Mounting type
    screw M4x22 (M4x27) 3 PCs., magnet (DVA141.121)
    stud M6x1 (M8x1/M10x1/M12x1) (DVA141.221)
  • Type of protection
    IP 54

Operational parameters

  • Operating temperature range, ºС
    -10…+80 (climatic version L)

Parameters of reliability and manufacturer's warranty

  • Mean time between failures, hours, not less than
    40 000
  • Warranty period, months
    18
  • Service life, years, not less than
    10
  • Calibration interval
    2 years